A sudden increase in CSF flow will close the ball and
cone valve and the entire volume of CSF will be forced
through the longer spiral passage, effectively slowing
the rate at which CSF is shunted from the brain. Once
the flow rate entering the SIPHONGUARD Device
decreases, the ruby ball separates from the valve seat,
opening the central passage. As long as CSF continues
to be shunted from the ventricles, flow through the spiral
passage of the SIPHONGUARD Device never stops,
regardless of the patient's position. See Graphs 1 and 2
for pressure flow characteristics.
Note: The SIPHONGUARD Device will not activate at
low CSF flow rates.
The SIPHONGUARD Device has a rigid enclosing shell
of polyethersulfone to prevent inadvertent closure
(and subsequent reduction or blockage of CSF flow)
caused by externally applied pressure.

Accessories

Accessory components supplied with some models of
the valve include:
Ventricular catheter
Distal catheter (unitized or separate)

Right angle adapter

Priming adapter

Catheters

The ventricular catheter is a 14 cm straight ventricular
catheter molded of radiopaque silicone elastomer with
x-ray detectable dots and a preassembled stainless steel
introducing stylet.
The distal catheter is 120 cm long, molded of radiopaque
silicone elastomer with x-ray detectable dots.
The BACTISEAL Catheters that are supplied with
some models of the valve are made of radiopaque
barium-impregnated silicone tubing that is subjected
to a process by which the silicone is impregnated with
rifampin and clindamycin hydrochloride.
The quantities of rifampin and clindamycin hydrochloride
used to impregnate the catheters are only a small
fraction of a therapeutic dose of these two antibiotics,
and have no potential for systemic therapeutic effect.
BACTISEAL Catheters are also available separately.
Right Angle Adapter
The right angle adapter, made of PROLENE
Material,
®
allows 90 degree bending of the ventricular catheter at
the burr hole site.
Priming Adapter
The priming adapter with inlet tubing facilitates pre-
implantation irrigation of the valve and catheters.

CODMAN CERTAS Tool Kit

Description

The CODMAN CERTAS Tool Kit is used to adjust and
confirm the performance setting of a CODMAN CERTAS
or CODMAN CERTAS Plus Programmable Valve. The
actions are performed both preoperatively (with the
valve still in the sterile package) and postoperatively
(on the implanted valve) through noninvasive means.
The CODMAN CERTAS Tool Kit employs magnetic force
to change the performance setting.
9
206521-001-G-INS.indd 9
21-05-2015 23:30:56