CLEANING, MAINTENANCE, AND STOR AGE
CLEANING
After each use, the TCU must be cleaned according to the following
procedure:
1. D i s c o n n e c t t h e T C U f r o m t h e P o w e r s u p p l y .
2. Wipe down the outside of the TCU housing with w arm w at er
using a sof t brist le brush unt il no soilage can be seen. DO
NOT SOAK OR IMMERSE.
3. Subsequently, wipe down the outside of the TCU housing with warm water
using a clean wipe.DO NOT SOAK OR IMMERSE
4. Disinfect the outside of the TCU housing with a clean wipe and
50%IPA/water solution. DO NOT SOAK OR IMMERSE
This unit contains electronic components. DO NOT soak, flush,
spray or use excessive free liquid on the TCU. Any units that have
been subjected to excessive free liquids during operation or cleaning
should not be used and must be returned to Idoman
MAINTENANCE
There are no user serviceable parts within the TCU. Opening the
device will void the warranty. Return unit to Idoman for service or
repair.
It is recommended that the TCU and its accompanying power suppl y are
subjected to annual inspection to ensure the safety of the device for both
patient and user.
This inspection should include the following:
Power connecting cords and power supply for signs of damage or wear.
TCU housing for signs of damage which may leave it in an unsafe or
contaminated state.
Presence and legibility of all Safety related markings and Labels.
Copy of Equipment Operators manual present.
Testing (If Required)
Protective earth resistance
Equipment leakage current
Patient leakage current
Insulation resistance.
Functional Test of Device
No modification of this equipment is allowed. Unauthorised modifications
or access may result in electrical shock or leave the unit in a hazardous
condition.
All TCU units that have reached their end of service life should be returned to
manufacturer for disposable as per WEEE Directive 2002/96/EC
STORAGE
1. Store the TCU in the Carrying Case. The storage temperature should
never exceed 50° C.
2. Store Disposable Cartridge in dry place at room temperature.
CAUTION
WARNING
14