GERATHERM Smart GT-1775 Manual De Instrucciones página 78

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 156
Appendix
NOTE:
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK
MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced
minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS
shall be calculated using the following equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
IMMUNITY TEST LEVEL in V/m.
EN
If it is necessary to achieve the IMMUNITY TEST LEVEL, the distance between
the transmitting antenna and the ME device or ME system may be reduced to 1 m.
The 1 m test distance is permitted by IEC 61000-4-3.
78
Tabla de contenido
loading

Tabla de contenido