Operating the Device
EN
Parameter
Description
P
User menu (accessible to the user)
L
Service menu (accessible only to service personnel)
P01
Volume level of pump unit alarm.
Two possible settings (1=muted, 2=loud).
P02
Setting, if balancing unit connected or not.
0 = not connected
1 = connected
first saved, then turn on again.
After saving the next menu points are not longer accessible.
P03
Setting for the deficit threshold - acoustic warning levels in the balancing unit
Setting range 0-2000 ml.
When pressing the key briefly -> in increments of 10
When pressing the key longer -> in increments of 100
Additional depiction in the deficit display
P04
Volume of the acoustic warning in the balancing unit
Two possible settings (1=muted, 2=loud).
Additional depiction in the deficit display.
P05
Check of the mass gauge of the balancing unit
Display in windows for deficit (0) – and nominal flow (0.00).
Nominal
Pressure
Nominal
Flow
28
5.13 Launching/Configuring the User Menu
Device parameters are changed and displayed with the user menu.
• Hold down the menu key (MENU) and turn on the device using the ON/OFF key.
• Keep the (MENU) key depressed until the LED test is finished. The nominal flow
display will depict a P after the LED test and emit an acoustic signal. The device
has now entered the first menu level of the menu mode.
• The first menu level offers a choice between User menu and Service menu. The
nominal pressure display depicts P for User menu or L for Service menu
(accessible only to service technicians).
• Switch from one to the other with the
• Confirm with the (MENU) key.
• Select the menu level or the value to be adjusted using the
• The values to be modified are displayed in the nominal flow display.
• Adjusted values are saved with the (MENU) key (use like ENTER key).
• Exit the menu or menu level without saving with the Start/Stop key
functioning as the ESC key.
• To reenter OP mode, turn the device off and on again.
The following user menu levels and settings are available:
/
keys.
/
keys.
Factory
Settings
2
1
750 ml
2
80 mmHg
150 ml/min