Step 3
To install a new radiator carefully take it out from the package - do not touch its transparent surface (in case
of any dirt - wash it with alcohol). Carefully slide the radiator into the housing, put on the joints, twist in the
grounding cord and turn on the device.
Reset:
7.5 RO module replacement
The RO module lifetime is 3-4 years. The lifetime of membrane cartridges may vary due to the flow rate, its
characteristics and level/types of feed water contamination.
• Symptoms of incorrect membrane functioning:
• Lower system efficiency, slower flow rate,
• Shorter lifetime of ion exchange columns.
To replace an RO module open the back door of the demineralizer. To do it, twist off four screws on the sides
and take the door out. Disconnect the RO module from the tubes and replace it with a new one. Close the
door and screw the screws.
67