General Information; How It Works; Mechanical-Sediment-Softening Filtracion - HYDROLAB HLP Serie Instrucciones De Uso

Desmineralizador
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

2.0 GENERAL INFORMATION

The demineralizer is fed by tap water. It produces water for analytical and instrumental purposes. Obtained
water fits the standards of the PN ISO 3696:1999, ASTM. Dependent on the demineralizer type, water may
be used for instrumental analyses: AAS, ICP/MS, IC, HPLC, GC. Demineralizers by default are equipped with
a microprocessor control and measurement system that supervises the demineralization process (type A au-
tomatics).

2.1 How it works

The demineralizer works under the tap water pressure. Purification processes are automated and mainte-
nance-free. These are:
1. Mechanical filtration
2. Sediment-carbon-softening filtration
3. Reverse osmosis (RO)
4. Ion exchange demineralization
5. UV lamp 254 nm*
6. Microfiltration capsule*
* depends on the model
2.2 Mechanical-sediment-softening filtration
5 μm pre-filter -prepares water for the reverse osmosis process. It contains a polypropy-
lene string cartridge wrapped around a filtration core located inside the filter. It filters out
all kind of mechanical filtration of size > 5 μm.
48