The demineralizer should be operating daily for at least 15 minutes.
Step 11*
General purpose water collection point.
1. Connecting an additional tank
If the device is equipped with an additional tank for III class water storage,
connect it to the socket "III TYPE 3696"
2. Connecting a washer, an autoclave or any other laboratory device.
- Connect the connector to the laboratory device.
- This connection has a built-in reduction on its end.
3. Installation of an additional tap
- If the point contains also a tap for water collection install it in the inox arm.
- Connect the tap cord to the socket "III TYPE 3696".
Amount of purified water stored in a pressure tank depends on a sum of tap water
and membrane pump pressure (if equipped). Differences in volume of stored water
Step 10
Connect the power cord to the socket "230 V"
Connect the power cord to the socket "24 V"
(HLP SMART)
CAUTION:
CAUTION:
may be up to 30%.
59