How To Replace A Microfiltration Capsule - HYDROLAB HLP Serie Instrucciones De Uso

Desmineralizador
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Step 1
- Turn off the device and close the feed water valve.
- Close the tank valve.
Step 2
Ion exchange module replacement:
- Open the front door and disconnect the exhausted ion exchange module.
- Connect a new module, mind the flow directions.
- Close the front door.
Step 3
Open the main valve and check if there are no leaks.
Step 4
The message „REPLACE MODULE H7/H6" disappears automatically,
when water conductivity drops below 4 μS/cm.
CAUTION!
3
After replacing the ion exchange resin first 10 dm
of produced water should be wasted.
7.3 Microfiltration capsule replacement
The lifetime of a microfiltration capsule is 12 months. The system displays a notification "REPLACE MF" and
triggers a sound alarm when the capsule needs to be replaced. Microfiltration capsule can be regenerated
during its operation. It is based on back-flushing and autoclaving the capsule in the max. temperature: 134 o C
(duration: 30 mins). This sterilization process is compatible with the ISO 11134 standard. A capsule may be
regenerated up to 3 times.
To change the capsule:
• Disconnect the exhausted capsule.
• During the replacement the tap must be closed.
• Connect a new capsule.
65