Discrete I/O

The discrete I/O interface options include both internal and remote I/O.
The internal version is available with dry-contact relay outputs. The relay
contacts will switch up to 30 volts DC or 250 volts AC. The inputs are switch
selectable as either active (for simple pushbutton control) or passive (for
connection to PLCs or other devices that supply their own power for the I/O).
The remote I/O is supported with the ARM100 remote module that provides dry-
contact outputs. The inputs are passive on the ARM100. An external 10 to 32
VDC power supply is required to operate the ARM100.
A total of 12 inputs and 18 outputs are supported through a maximum of three
options.

Ethernet/COM2/COM3 Option

The Ethernet port can be used for FTP transfer of tare and target tables and
complete setup files. It also provides a TCP/IP port to transmit a demand template,
continuous data, for backup and restore of the terminal's configuration using the
METTLER TOLEDO InSite
server, and to send email alerts when calibration expires or fails.
COM2 provides RS-232 communication at rates from 300 to 115.2k baud. COM 3
supports the same baud rates and provides an RS-232, RS-422, or RS-485
connection.
1-12
EtherNet/IP™
PROFIBUS
DP
SL Configuration Tool
THE INTERNAL #71209093 DISCRETE I/O RELAY OPTION MUST NOT BE
USED IN AN IND560 TERMINAL INSTALLED IN AN AREA CLASSIFIED AS
DIVISION 2 OR ZONE 2/22. FAILURE TO COMPLY WITH THIS WARNING
COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE.
SL program, for direct access to data via a shared data
Modbus TCP
WARNING