G
Manual leg spreader handle (1 piece, optional)
H
Sling (1 piece, optional)
1.4 Service Life
WARNING!
Risk of injury
Maintenance MUST be performed only by qualified
personnel.
Improper assembly may cause injury or damage.
– Regular maintenance of patient lifts and accessories is
necessary to assure proper operation.
– DO NOT overtighten the mounting hardware. This
will damage the mounting bracket.
The mobile patient lift has an expected lifetime of 8 years when used
in accordance with safety instructions, maintenance intervals and
correct use, stated in this manual. The effective product lifetime can
vary according to frequency and intensity of use.
Actuator Service Life
Number of lifts per day
1–2
3
4
5
6
1575480-A
Service life of actuator
(in years)
10
9
6
5
4
Number of lifts per day
7
10–13
14–27
1.5 Warranty Information
Terms and conditions of the warranty are part of the general terms
and conditions particular to the individual countries in which this
product is sold.
Contact information for your local Invacare office is located inside
the back cover of this manual.
General
Service life of actuator
(in years)
3
2
1
7