Oxygen Concentration vs Flow Rate (Across the listed voltage and environmental conditions.)
525DS, 525DS-Q
Flow L/m
%O
2
5
87% - 96%
4
87% - 96%
3
87% - 96%
2
87% - 96%
1
87% - 96%
.5
87% - 96%
ELECTROMAGNETIC COMPATIBILITY INFORMATION
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the oxygen
concentrator or MR medical devices . The device and accessories have not been evaluated for safety in an MR environment .
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security
systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and
discontinue use . Contact your home care provider .
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high .
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation . If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating normally .
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the oxygen concentrator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
EN - 12
525KS, 525KS-LT, 525PS
Flow L/m
%O
2
5
90% - 96%
4
90% - 96%
3
90% - 96%
2
90% - 96%
1
90% - 96%
.5
90% - 96%
SE-525K4