6.4.2
UNIT-MOUNTED MASTER FLUX PLUS
Screw the hoses (Figure 2 Pos 8/9) to the unit-mounted.
Connect the mask circuit gas exhaust tube to the supplied fitting (Figura 17).
Connect the suction circuit (Figure 16) to the hose fitting.
Connect the fitting to the surgical aspirator tube (Ø11 mm).
WALL-MOUNTED MASTER FLUX PLUS
Remove the tube (Figure 16 Pos 43).
Connect the tube, gas exhaust of the mask circuit, to the valve (Figure 16 Pos 44).
Connect the fitting to the surgical aspirator tube (Ø11).
VALVE ADJUSTMENT
Turn the perforated ring nut on the valve (Figura 16 Pos 44) until the holes close.
Start gas delivery by setting ~10 l/min.
Start the surgical aspirator by setting the minimum aspiration value.
Observe the bar (Figura 13 Pos F).
Turn the perforated ring nut on the valve gradually opening the holes until the bag reaches
its ideal condition of "HALF FULL HALF EMPTY"
By changing the flow delivery it may be necessary to adjust the correct aspiration
flows of the exhaled gases.