General Problems - Tecno-gaz Multisteril Fast Instrucciones De Uso

Ocultar thumbs Ver también para Multisteril Fast:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 108
Display

10.4 General problems

Problems
Poor cleaning
Poor drying
Some instruments
get damaged
Causes
Water infeed alarm
Water infeed alarm
Pressure of the water
system not normal
TANK WATER DISCHARGE
IT STARTS THE DISCHARGE PUMP FOR 30 SECONDS
Possible causes
- Calcareous water, white halos on
the tools
- Traces of elements which have
not been pre-treated (cements,
alginates)
- Wrong cycle parameters
- The ultrasound noise is not heard
during the cleaning phase
- Load excessive, drops on tools
- Wrong installation, rests of water
on the tank bottom after the
draining.
- No air flow from the lid side slots
- Instruments that cannot be treated
with ultrasounds cleaning (e.g.
Mirrors)
Solutions
PROBE J - Y
Clean the metal with products suitable for
stainless steel or a stripper.
Use a sponge (slightly abrasive).
PROBE J - Y
Clean the metal with products suitable for
stainless steel or a stripper. Use a sponge
(slightly abrasive).
Make sure water pressure ranges are
within the indicated values. If the pressure
is insufficient, contact a plumber.
Keep pressed for 8s. button OK
Solution
- Install a water softener before the
device.
- Some detergents might reduce such
effect
- Pre-treat with suitable detergents.
- Modify parameters increasing the
cleaning time .
- Contact the technical service
- Reduce the load
- Position the device correctly, modify
the height of the feet
- Contact the technical service
- Do not clean with ultrasounds
instruments that cannot be treated
with ultrasounds treatment