4.8 Storing splicing results

4.5.1 The splicing result is automatically stored in memory when
the [Finish] screen, or when the wind protector is opened
upon completion of the splice at the [Finish] screen.
4.5.2 After the 4000th result is stored, 4001st splice result is
written over 1

4.9 Reinforce the fusion point

<1> Open tube heater lids and transfer fiber with protection sleeve
from V-grooves of the Fusion Splicer to tube heater.
<2> Make sure the splice point is located at the center of the
protection s leeve. Put the protection sleeve at the center of
theheater. Apply some tension on the fiber so the tubeheater
lids close.
Notes:
z Make sure the strength member in the protection sleeve is placed
downwards
z Ensure no retorting of the fiber.
<3> Press
to start tube heating. The heating process includes: start
heating (the LED indicator turns green); constant temperature
period (the LED indicator turns red); temperature fall period (the
LED indicator turns green); the buzzer beeps and the HEAT LED
Indicator (green) turn off when heating is completed. If
pressed again, the tube heating process is aborted.
<4> Open tube heater lids and remove protected fiber from the tube
heater.
z Protection sleeve may stick to bottom plate of heater. Use a cotton
swab to help remove sleeve from heater.
<5> Visually inspect the finished sleeve to verify no bubbles or
debris/dust is present in the sleeve.

4.7 Manually move the motor

The four motors (forwarding, adjusting the core)
in the Fusion Splicer can be manually
operated individually. In the "edit splice operation"
menu, set [pause 1] and [pause 2] to "ON".
The motor will automatically paused when
the fusion splice process reaches the stage
of [pause 1] and [pause 2]. At that time,
©Copyright Abacanto Digital SA.
or
is pressed upon completion of the splice at
st
.
Manual de usuario AD300
AD Instruments
is