Full HD 3CCD Camera system PV440
Malfunction
Color bar displayed on the monitor
Inaccurate color representation on the monitor
8.1

Fuse replacement

Risk of fatal injury from electric shock!
Unplug the device before changing the fuses!
DANGER
Specified fuses: T 0.5 A L/250 V~; Art. no. TA020394
Use a small screwdriver to release the clip on the fuse holder.
Pull out the fuse holder.
Replace both fuses.
Reinsert the fuse holder in such a way that it audibly snaps into place.
Fig. 3
Fuse replacement
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
14
®
Cause
Camera cable not connected to the camera con-
trol unit
White balance failed
Inappropriate monitor color settings
Video cable defective
Remedy
Connect camera cable at the camera control
unit
Initiate white-adjustment
Reset to factory setting
Replace video cable
9.

Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
For service and repairs, please contact your national B. Braun/
Aesculapagency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Repairs handling camera systems:
Aesculap FLEXIMED GmbH
Repair Service
Robert-Bosch-Str. 1
79211 Denzlingen/Germany
Phone:
+49 7666 93 21 0
Fax:
+49 7666 93 21 580
E-mail:
Telephone hotline for technical assistance with electrical devices and
device combinations:
Phone:
+49 7461 95 1601
Other service addresses can be obtained from the address indicated above.

10. Accessories/Spare parts

Art. no.
Designation
PV442
Camera head and camera cable
PV441
Camera cable
PV126S
Endo-lens with zoom
PV127S
Fixed focal length endo-lens
TA020394
Fuse set T 0.5 A L/250 V~