• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
• When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
8. Rectifying faults
Error messages can occur when
• you move or talk during the measurement (
• the cuff tube is not properly inserted (
• when changing the batteries, the start/stop button
pressed,
• inflation takes longer than 20 seconds (
• the inflation pressure is higher than 300 mmHg (
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.
9. Specifications
Model no.
BM 60
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 40 –180 mmHg,
Pulse 40 –160 beats/minute
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
,
),
Cuff size
),
was
Permissible operating
conditions
),
).
Permissible storage
and transport con-
ditions
Power supply
Battery life
Accessories
Classification
Maximum surface
temperature
23
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 101 mm x W 130 mm x H 54 mm
Approx. 387 g (without batteries)
22 to 32 cm
+10 °C to + 40 °C, ≤ 85 % relative
air humidity (non-condensing),
700 –1060 hPa ambient pressure
- 5 °C to + 50 °C, ≤ 85 % relative air
humidity
4 x 1,5 V
AA batteries
For approx. 300 measurements,
depending on the blood pressure level
and/or pump pressure
Instruction for use, 4 x 1.5 V AA batter-
ies, storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF (cuff), CLASS II (when using a
mains part)
Device: 44,1 °C
Mains part: 47,2 °C