Initiating a Single NIBP Measurement
To take a single NIBP reading:
Step
Action
1
Enter a new patient ID or select a previously saved patient ID. If you choose not to enter an ID, the
measurements will be assigned to an "Unknown" patient ID.
2
Verify that the correct NIBP mode is selected: Adult/Pediatric or Neonatal.
3
Choose the appropriate cuff and place it on the patient.
4
Press the Cuff Start/Stop button. When the measurement is complete, the Systolic, Diastolic, and Mean
Arterial Pressure values appear in their corresponding LEDs.
Continuous NIBP Monitoring
The VS1 monitor can be used to continually measure a patient's blood pressure at specified intervals. You
can also create a cuff interval program that measures blood pressure at predetermined times and in
predefined intervals.
In Adult mode, the VS1 monitor will sound an alarm at a systolic pressure below 70 mmHg (9.3
Note—
kPa), even if a lower limit is set. Neither the feature nor the alarm can be overridden. The range remains
adjustable for use in the Neonatal mode, where lower limits can be set.
Selecting a Cuff Interval
The Cuff Interval option initiates a blood pressure reading every n minutes. From the Cuff Interval screen,
you can select any of the following intervals: OFF, STAT, 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, or180
minutes.
CURRENT ID:
MIKE WALLACE, 123458
PREV
SCREEN
If you select an interval of 1 minute, after 12 measurements, the monitor automatically changes the
Note—
interval measurement to 5 minutes.
To change the cuff interval value:
Step
Action
1
From the Line List display, press the Cuff soft key. The Cuff Interval screen appears. The currently
selected cuff interval is displayed.
INTERVAL: 2 MIN
STOP
PROGRAM
Initiating a Single NIBP Measurement
11:06
STAT
SAVE
->MAIN
Monitoring Blood Pressure
5-5