► Remove the adapter cover before the patient leaves the workshop/fit­
ting area.

5.1.1 Applying/removing the footshell

CAUTION
Using the prosthetic foot without footshell
Risk of injury due to lack of surface grip and loss of functionality of the
prosthetic foot
► Always use the prosthetic foot with a suitable footshell.
► Apply or remove the footshell as described in the footshell instructions
for use.

5.1.2 Bench Alignment

TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring
apparatus, 743A80 50:50 gauge, alignment apparatus (e.g. 743L200
L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Assemble and align the prosthetic components in the alignment apparatus
according to the following specifications:
Heel height: Effective heel height (shoe heel height - sole thick­
   
ness of forefoot) + 5 mm
Sagittal alignment: Alignment reference line on a–p markings
   
of the connection cap/footshell (see fig. 2)
Frontal alignment: Alignment reference line on m–l markings of
   
the connection cap/footshell (see fig. 2)
   
Connect the prosthetic foot and prosthetic socket using the chosen
adapters. Be sure to follow the adapter instructions for use.
   
Sagittal alignment:
Determine the centre of the prosthetic socket with the 50:50
gauge. Align the prosthetic socket centrally to the alignment refer­
ence line.
Socket flexion: Individual residual limb flexion + 5°
   
Observe the abduction or adduction position.
16
Bench alignment process