Specifications
Display
LC display, 7-segment with icons
Main display
Character height 17 mm, unit symbols 10 mm
Secondary display
Character height 10 mm, unit symbols 7 mm
Sensoface
3 status indicators (friendly, neutral, sad Sensoface)
Mode indicators
5 status bars "meas", "cal", "alarm", "cleaning",
"config"
18 further icons for configuration and messages
Alarm indication
Red LED in case of alarm or HOLD, user defined
[cal] [conf] [ ] [ ] [enter]
Keypad
5 keys:
Service functions
Current source
Current specifiable for output 1 and 2 (00.00 to 22.00mA)
Manual controller
Controller output entered directly (start of control process)
Device self-test
Automatic memory test (RAM, FLASH, EEPROM)
Display test
Display of all segments
Last Error
Display of last error occurred
Sensor monitor
Display of direct sensor signal (resistance/temperature)
Relay test
Manual control of the four switching contacts
Parameter sets
*)
Two selectable parameter sets for different
process phases
Switchover via CONTROL input or manually
Signaling via relay contact PSEt2
Data retention
Parameters and calibration data > 10 years (EEPROM)
102
Protection against Protective separation of all extra-low-voltage circuits against
electrical shock
Power supply
Nominal operating conditions
Ambient temperature
Transport/Storage temp -20 to +70 °C
Relative humidity
Power supply
Frequency for AC
EMC
Emitted interference
Immunity to
interference
Explosion protection
FM:
CSA:
Enclosure
Color
Assembly
Dimensions
Ingress protection
Cable glands
Weight
*) User-defined
1) To IEC 746 Part 1, at nominal operating conditions
2) ± 1 count
3) Plus sensor error
Cond Ind 7100 e
mains by double insulation as per EN 61010-1
24 (-15%) to 230 V AC/DC (+10%); approx. 5 VA, 2.5 W
AC: 45 to 65 Hz; Overvoltage category II, Class II
-20 to +55 °C
10 to 95 % not condensing
24 (-15%) to 230 V AC/DC (+10%)
45 to 65 Hz
EN 61326
Class B (residential environment)
Class A for mains supply > 60 V DC
Industrial environment
NI Class I Div 2 Group A, B, C & D, T4
Ta = 55 °C; Type 2
NI Class I Zone 2 Group IIC, T4 Ta = 55°C; Type 2
Class I Div 2 Groups A, B, C and D, T4
Ex nA IIC T4
molded enclosure made of PBT (polybutylene terephtalate)
Bluish gray RAL 7031
• Wall mounting
• Pipe mounting: dia 40 to 60 mm,
30 to 45 mm
• Panel mounting, cutout to DIN 43 700
Sealed against panel
H 144 mm, B 144 mm, T 105 mm
IP 65 / NEMA 4X
3 breakthroughs for cable glands M20x1.5
2 breakthroughs for NPT 1/2 " or
Rigid Metallic Conduit
Approx. 1 kg
103