Invacare MicroAIR MA50CE Serie Manual Del Usuario página 21

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 94
Firm/Soft Buttons
1.
Select comfort pressure settings by pressing Firm (
Soft Button - Pressing this button reduces the pressure setting in the mattress.
Firm Button - Pressing this button increase the pressure setting in the mattress.
The patient comfort pressure ranges from Soft (level 0 = 6±4 mmHg) to Firm (level 9 = 32±6 mmHg). The Comfort Control LED
displays the patient comfort pressure levels from 0 to 9 and provides a guide to the caregiver to set approximate comfort pressure level
depending on the patient weight. If the patient's weight to height ratio is above average, increase the pressure setting by approximately
20%.
Fowler Button
Use when articulating the bed to a sitting position by raising the head.
1.
Press the Fowler button (
The Fowler LED illuminates when in this mode. When this mode is activated, the control unit increases the pressure in the overlay to
prevent the patient from bottoming out.
Max Inflate Button
1.
Press the Max Inflate button (
± 6 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 20 minutes, the Max Inflate
mode deactivates and the control unit defaults to the previous setting.
It takes 20-45 minutes for the overlay to inflate fully (inflation time depends on size of overlay).
Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED.
It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning.
Part No 1176084
) or Soft (
) to activate the patient fowler mode.
) to select the Max Inflate mode and inflate the overlay rapidly to maximum firmness (pressurized to 35
) button.
21
4 OPERATION
Invacare® MA50CE Series
Tabla de contenido
loading

Tabla de contenido