Adult Respiratory Sensor
SpO
, RR
2
Identification of a substance that
is contained or present within the
product or packaging
Not made with
natural rubber latex
Identification of a substance that is
not contained or present within the
product or packaging
Protected against access to hazardous
22
parts with a finger and protected
against vertically falling water drops
when sensor tilted up to 15°
This product cannot be adequately cleaned and/or sterilized by the
user in order to facilitate safe reuse, and is therefore intended for single
use. Attempts to clean or sterilize these devices may result in bio-
incompatibility, infection or product failure risks to the patient.
This device is not made with natural rubber latex or DEHP.
Directions for Use
Indications/Contraindications
The Nellcor™ Adult Respiratory Sensor, when used in conjunction with
a Nellcor pulse oximeter and the Nellcor Respiration Rate Software, is
indicated for single patient use when continuous noninvasive arterial
oxygen saturation, pulse rate and respiration rate monitoring are required
for adult patients weighing more than 30 kg.
The Nellcor Adult Respiratory Sensor is contraindicated for use on patients
who exhibit allergic reactions to the adhesive tape.
Instructions for Use
1. Remove plastic backing from the sensor and locate transparent
windows (a) on the adhesive side. Windows cover the sensor's optical
components. (1)
An index finger is the preferred sensor location. Alternatively, apply the
sensor to a small thumb, small finger, or great toe.
Note: When selecting a sensor site priority should be given to an extremity
free of an arterial catheter, blood pressure cuff, or intravascular
infusion line.
Do not use if
package is opened or
damaged
Not made
with DEHP
Do Not Use
with MRI
3