>FUNCTION IN1
▼▲ ME
*
NORMALLY OPEN CONTACT
Normally closed contact
▼▲
>
TYPE OF WATER METER
ME
1 LITRE/IMPULSE
2,5 Litres/Impulse
5 Litres/Impulse
10 Litres/Impulse
100 Litres/Impulse
*
500 Litres/Impulse
1000 Litres/Impulse
32
Function IN1
An external measuring point switch-over takes place via this input if
you have programmed "
status of IN1 has to be programmed according to the output function
of the controller (volt-free contact required!).
In
, select
>BASIC PROGRAM
=>
PROGRAM VALUES
POINTS EXT.
Press "ENTER" to confirm the entry
In
, select
>BASIC PROGRAM
=>
PROGRAM VALUES
=>
Select a normally closed or normally open contact
Press "ENTER" to confirm the entry
Water meter
It is necessary to connect a water meter to input IN2 for quantity-
dependent analysis start. Program the corresponding water meter
rating.
In
, select
>BASIC MENU
=>
=>
PROGRAM VALUES
Select the water meter rating
Press "ENTER" to confirm the entry
Entering further basic program data
2 measuring points externally
=>
=>
MEASURING POINTS
FUNCTION IN1
TYPE OF WATER METER
". The active
2 MEAS.