Extracting the reagent
Switch the instrument on and press the "STANDBY" key.
During operation, the pump (PERIClip)
To ensure that reagent is available for the initial analyses, the
Press the "manual"
If necessary, manually tighten the hose connectors of the intake
Input reagent level
Opening the water inlet
If water sprays from the hose
Instrument settings and data input
Please read the following information before carrying out settings and
entering data for operating the instrument.
reagent.
intake hose
and the transport hose
indicator from the pump up to the measuring chamber.
and the transport hose
and transport hose slightly in case of bubble formation.
In the program menu M select
>Service I -> Input Reagents
Set the reagent levels to 100%.
Press ENTER to confirm the selection.
Open the lower housing cover.
Slowly open the manually operated shut-off valve to prevent the
measuring chamber overflowing. The flow regulator requires a few
seconds to function correctly.
Make sure that the water conducting parts are not leaky.
reduce the amount of inlet water via the manually operated shut-off
valve. It should take 2 to 6 seconds to fill the measuring chamber!
automatically extracts
must be filled with
key several times until the intake hose
are filled up to the measuring chamber.
of the measuring chamber
,