Alarm/Message; Function Aux - Heyl Testomat 2000 PO4 Manual De Instrucciones

Fotómetro de procesos para fosfato 0 –10 mg/l
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

>ALARM/MESSAGE
▼▲ ME
A/M/-
REAGENT LOW LEVEL
A
Low water pressure
A
A/M
A/M
Ff. optics
A
Ff. dosing pump
A
A/M
A/M
Ff. outlet to drain
M
A/M/-
Mf. dirtiness
A
Power failure 24V
M
A/M
A/M/-
Mf. turbid
M
Meas. range exceeded M
A/M/-
A/M/-
Maint. int. exceeded
M
Change pump head 1
A
A/M
A/M
Change pump head 2
A
A=Alarm,
M=Message
- = no action
Ff.=Function fault
Mf.=Measuring fault
NOTE
>FUNCTION AUX
▼▲ ME
CONTACT BEFORE ANALYS.
*
*
Contact during analysis
Contact „Suction"
Time:
00m:10s

Alarm/Message

The instrument is equipped with an alarm relay output for signalling
faults. The events which mean a fault at the instrument or are
intended to trigger a message, can either trigger an alarm "A"
(continuous contact) or a message "M" (2-second impulse).
The faults are recorded and stored in the error history if the event has
been programmed as an alarm or message. For example, if a low
reagent level has not been programmed as an ALARM/MESSAGE, it
is not registered in the error history. Up to 20 error messages can be
stored. A list of these errors can be opened in the information menu.
The information stored per event is the time (day, month, year and
hour) and the type of the error.
 In >
, select
BASIC PROGRAM
=>
=>
PROGRAM VALUES
 Select the type of monitoring A=alarm, M=message or - = no action
for the individual menu items
 Press "ENTER" to confirm the respective entry
Error messages
 All error messages are lost after a power failure.
 Certain instrument faults always trigger an alarm or a message
(see chapter "Description of the relay outputs")!

Function AUX

The AUX relay output can be programmed for the following control
functions:
As a function output for the contact with programmable duration
prior to and/or during the analysis, or after an analysis.
For example, it is possible to control the cooling water inlet of an
upstream cooler via a solenoid valve. This ensures that the cooling
water only flows when required, i.e. when an analysis is being carried
out.
 In >
, select
BASIC PROGRAM
=>
PROGRAM VALUES
=>
 Select the program step at which the AUX contact is to be
activated
 Under "Time" enter the contact duration in minutes (m) and
seconds (s)
 Press "ENTER" to confirm the entry
Entering further basic program data
ALARM/MESSAGE
FUNCTION AUX
33