Error Messages/Troubleshooting - Heyl Testomat 2000 PO4 Manual De Instrucciones

Fotómetro de procesos para fosfato 0 –10 mg/l
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Displayed message (flashes at
selected display)
Ff. POWER FAILURE 24 V
CANCEL WITH HORN KEY
Ff. DOSING PUMP
CANCEL WITH HORN KEY
Mf. TURBID
CANCEL WITH HORN KEY
MEASURING RANGE EXCEEDED
CANCEL WITH HORN KEY
LOW WATER LEVEL
CANCEL WITH HORN KEY
Ff. OUTLET TO DRAIN
CANCEL WITH HORN KEY
LOW REAGENT LEVEL
CANCEL WITH HORN KEY
Mf. SOILING
CANCEL WITH HORN KEY
Ff. OPTICS
CANCEL WITH HORN KEY
MAINTENANCE INTERVAL
EXCEEDED BY
XXX DAYS
CANCEL WITH HORN KEY
CHANGE PUMP HEAD 1
CHANGE PUMP HEAD 2
Abbreviations: Ff.: = function fault, Mf. = Measuring fault

Error messages/Troubleshooting

Instrument result functions
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses or no message
- Continue measurements
- After programming:
Continuous alarm or message
impulses or no message
- Continue measurements
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses or no message
- LED and output
"maintenance" on
- Continue measurements
- After programming:
Continuous alarm or message
impulses or no message LED
and output "maintenance" on
- Continue measurements
- After programming:
Continuous alarm or message
impulses
- Standby
- After programming:
Continuous alarm or message
impulses or no message
- LED and output
"maintenance" on
- Continue measurements
- After programming:
Continuous alarm or message
impulses or no message
- LED and output
"maintenance" on
- Continue measurements
Error messages/Troubleshooting
Possible causes
- Internal power failure of the
24 V supply
- Dosing pump is defective
- No dosing message from
the dosing pump
- The water is turbid / soiled
- The measuring range has
been exceeded
- No water input although
LED "IN" lamp illuminates
- Inlet pressure too low
- Overflow detection does not
react
- Water remains in the
measuring chamber
although LED "OUT"
illuminates
- Minimum reagent quantity
not reached
- Sight-glass windows are
soiled
- Plug-in circuit board
defective
- Error at the optical
component (transmitter or
receiver defective)
- Programmed maintenance
date reached or exceeded
- end of lifetime
Remedies
 Replace fuse F4 or F8
(The control lamp "Power"
of the dosing pump should
illuminate)
 Replace dosing pump
 Check cable to the dosing
pump for correct
connection
 Check water inlet
 Connector at the inlet
valve oxidised
 Clean filter strainer
 Replace valve block
Extract pressure controller
valve body
 Replace fuse F6
 Check water outlet
 Connector at the outlet
valve oxidised
 Replace valve block
 Check reagent level and,
if necessary, refill
(enter the filling quantity!)
 Clean sight-glass windows
 Replace plug-in circuit
board
 Replace the measuring
chamber holder
 Carry out maintenance
and subsequently cancel
or confirm
 Change pump head and
confirm
47