WARNING
NOTE
8
Installation
Always completely disconnect the relevant plant part before
installing the instrument or connecting/disconnecting it to/from the
power supply. Secure the plant against reconnection.
Only connect the instrument to the mains voltage specified on the
rating plate.
Always observe technical data and ambient parameters.
Testomat 2000
®
PO4 requires an interference free and stable
power supply. If necessary, use a mains filter to protect Testomat
®
2000
PO4 against interference voltages caused, e.g., by solenoid
valves or large motors. Never lay connecting cables parallel to
power cables.
Operation
Ensure that the maximum electrical load capacity of the relay
outputs is never exceeded.
Immediately switch off Testomat 2000
staff if malfunctioning occurs. The warranty will be void if you
tamper with or attempt to repair Testomat 2000
must be carried out by authorised service staff.
Cleaning
Only use a dry, lint-free cloth for cleaning.
After switch-off and longer downtime
Ventilate the indicator leads as described in the start-up, because
longer downtimes (more than 6 hours) can cause the indicator in
the leads to retract.
Do not switch off the appliance for longer periods (e.g. over the
weekend) via the start/stop output. The indicator can retract from
the leads. This results in measurement errors after the appliance is
switched on.
De-installation
Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a
defective instrument (irrespective of the warranty period) if it has
been de-installed and returned to us with a description of the error.
Disposal
Dispose of the instrument in accordance with national regulations.
Important safety information
®
PO4 and contact service
®
PO4. Repairs