Password Protection And Basic Program; Selecting The Operating Mode - Heyl Testomat 2000 PO4 Manual De Instrucciones

Fotómetro de procesos para fosfato 0 –10 mg/l
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Testomat instrument
(display) in display mode
▼▲
>PROGRAM
ME
SERVICE I
Basic program
Date/Time of day
BASIC PROGRAM
▼▲
>BASIC PROGRAM
ME
09.01.09
10:25
Password:
0000
▼▲
>BASIC PROGRAM
ME
PROGRAM VALUES
Service II
Customer service
Time controlled
Internal start via a
timer.
▼▲
>
ME
MODE OF OPERATION
TIME CONTROLLED
*
Volume interval
Dynamics
External (Start)
24
Password protection and basic
program
A four-digit password is required to enter data and to carry out
settings in the basic program. The password is the current time of
®
Testomat 2000
in reverse order.
Password entry
 Press the "M" key
The basic menu "
 Use the cursor block
"
"
Basic program
The selection appears in CAPITAL LETTERS.
 Press "ENTER" to confirm your selection
The selected submenu "
 Press "ENTER" to confirm the menu item "
The cursor flashes in the
 Use the cursor keys
"5201"
 Press "ENTER" to confirm the entry
The selection menu for the basic program appears. You can now
enter the plant specific data.
Please enter the
time in reverse
order:
=> 5201

Selecting the operating mode

Under the menu item
type of analysis controller. Testomat 2000
selection options: Time control, quantity control via water meter and
external analysis start.
Shortest interval = 0 minutes between analyses. Largest interval =
255 minutes.
The analysis interval (time between two analyses) is determined by
the duration of the supplementary program AUX, the set flush times
(internal and external), the programmed interval and the duration of
the analysis. The analysis duration depends directly on the
measured value.
Selecting the time control
 In
>BASIC PROGRAM
=>
PROGRAM VALUES
CONTROLLED
 Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
("TIME CONTROLLED" " * " is preset)

Password protection and basic program

" appears
>PROGRAM
to select the desired menu item
>BASIC PROGRAM
" appears
"Password:" field
to enter the time in reverse order:
"Mode of Operation"
®
, select
=>
MODE OF OPERATION
"
BASIC PROGRAM
█000
it is possible to select the
PO4 provides numerous
=>
TIME
Tabla de contenido
loading