Carrying out a bump test

A bump test must be carried out on every device
before use.
The bump test can be carried out in two ways.
Bump test for the Dräger X-am 5000:
z Perform a bump test in accordance with the
Instructions for Use for the gas detection instrument
before using with the Dräger X-zone 5000.
Bump test on the Dräger X-am 5000 in combination
with the Dräger X-zone 5000:
NOTICE
The bump test can only be carried out with the Dräger
X-am 5000 holder diffusion mode
(order no. 83 20 636).
z Switch the Dräger X-zone 5000 on (refer to
"Switching on the instrument" on page 29).
z Place adapter (1) (order no. 83 20 108) on the holder
(diffusion).
z Connect the test gas cylinder to the adapter (2).
z Open the test gas cylinder valve to let test gas flow
over the sensors.
z Wait until the instrument displays the test gas
concentration with sufficient tolerance:
1)
Ex: ±20 %
1)
O
: ±0.8 Vol.-%
2
1)
TOX: ±20 %
.
– Depending on the test gas concentration, alarm A1
or A2 is issued when the alarm thresholds are
exceeded.
z Close the test gas cylinder valve and remove the
adapter from the holder.
If the displays are outside of the above-mentioned
ranges: Have the Dräger X-am 5000 calibrated by
service personnel.
1) Upon application of the Dräger mixed gas (order no. 68 11 130)
the displays should be within this range. Different concentrations
can be set using the Dräger CC-Vision PC software provided.
CAUTION

Carrying out a connection test

The connection test can be used to check the correct
connection to all Dräger X-zone 5000 devices.
z Press the
key on one of the the Dräger X-am 5000
OK
devices being used three times in succession.
A signal via horn and LED ring is issued three times on
every device connected wirelessly or via a cable.

Safety housing

1
2
In exceptionally rough environments, and in very strong
sunlight and high temperatures (> +40 °C), the use of
the safety housing (order no. 83 21 519) is
recommended.
1 Hose/cable holder

Alarm-attenuation ring

The alarm-attenuation ring must not be used in
explosion-hazard areas!
When switching on and off and during the bump test of
the Dräger X-zone 5000, Dräger recommends wearing
hearing protection or using the alarm-attenuation ring
(order no. 83 20 110) as the acoustic alarm is activated
for a short time.