3. Specifications
Specifications
Long-term stability
Permissible temperature ranges
GB
Medium
5) 6)
■
Ambient
5)
■
Storage
5)
■
Compensated temperature range
Temperature coefficients in the
compensated temperature range
Mean TC of zero point
■
Mean TC of span
■
CE conformity
Pressure equipment directive
■
EMC directive
■
Shock resistance
Vibration resistance
Reverse polarity protection
Weight
{ } Items in curved brackets are optional extras for an additional price.
1) Applies only to model D-10-7.
2) For model D-11-7: The value specified in the table applies only when sealing is made using a sealing ring below the hexagon. Otherwise
max. 1,500 bar applies.
3) Not with model D-10-7 for measuring ranges > 25 bar.
4) Including non-linearity, hysteresis, zero offset and end value deviation (corresponds to measured error per IEC 61298-2). Calibrated in
vertical mounting position with process connection facing downwards.
5) Also meets EN 50178, tab. 7, operation (C) 4K4H, storage (D) 1K4, transport (E) 2K3
6) Model D-11-7 is not available in an oxygen version. Model D-10-7 is only available in an oxygen version with medium temperatures
between -20 ... +60 °C.
For special model numbers, e.g. D-10000-7 or D-11000-7, please note the specifications stated on the
delivery note.
For further specifications see WIKA data sheet PE 81.30 and the order documentation.
10
Models D-1x-7
% of span
≤ 0.10 / year (at reference conditions)
°C
-20 ... +80
°C
-20 ... +80
°C
-40 ... +85
°C
-20 ... +80
(The temperature errors in the range 0 ... +50 °C are already
included in the accuracy)
% of span
≤ 0.20 / 10 K {≤ 0.10 / 10 K}
% of span
≤ 0.20 / 10 K {≤ 0.10 / 10 K}
97/23/EC
2004/108/EC, EN 61326 emission (group 1, class B) and
interference immunity (industrial application)
g
< 100 per IEC 60068-2-27 (mechanical shock)
g
< 5 per IEC 60068-2-6 (vibration under resonance)
U
vs. U
B+
kg
approx. 0.4
WIKA operating instructions pressure transmitter model D-1x-7
B-