WIKA D-1-7 Serie Manual De Instrucciones página 21

Transmisor de presión
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
Zero adjustment
Output data
0x21
XX
0x23
HI
The respective correction value can be calculated as follows:
Correction value = -
Correction value: Digit value which is programmed into the instrument via service 0x23.
Offset: Deviation of the zero point from the set point in bar
Span: MBE - MBA in bar
Changes only take effect after a reset.
Zero adjustment, example 1:
A pressure transmitter with a measuring range of -1 ... +2.5 bar has a zero offset of +10 mbar, then one
would correct:
Measuring range -1 ... +2.5 bar
Display value 10 mbar too high
Correction value = -
0.010 bar
Correction value = -
WIKA operating instructions pressure transmitter model D-1x-7
Input data
XX
0x21
LO
0x23
Offset
x 50,000
Span
Span: 3.5 bar
Offset: +0.010 bar
Offset
x 50,000
Span
x 50,000
3.5 bar
HI
LO
HI
LO
Correction value = -143
Explanation
Read zero adjustment
Write zero adjustment
0xFF71
Output data
0x23
0xFF
GB
0x71
21
Tabla de contenido
loading

Tabla de contenido