6.0
Fig. 6. Water filter
Filter body
Filter cartridge
Sealing ring
Plexiglass cup
44
6.5
Replacement of filter
When the filter cartridge is very dirty (dark colour) the filter
cartridge is to be exchanged as follows:
• Close water tap of supply.
• Switch on unit to reduce pressure in the filter body.
• Switch off unit after appearance of "F 1".
• Unscrew Plexiglass cup (
(
, fig. 6) by turning anti-clockwise.
• Extract filter cartridge (
one (Art.No. 502.0891.0).
The filter element must not come into contact with
contaminated objects in order to prevent the ingress of
germs into the water system.
UK
, fig. 6) from filter body
, fig. 6) and replace by new