Instrument Self-Test - Mettler Toledo SevenGo Duo pro SG78 Instrucciones De Manejo

Ph/orp/ion/conductivity meter
Ocultar thumbs Ver también para SevenGo Duo pro SG78:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Setup

5.7 Instrument self-test

The instrument self-test requires user interaction.
a) In the Setup menu, select "6. Instrument Self-test".
b) Press the function keys on the keypad one by one in any order.
Notes
The user needs to finish pressing all seven keys within two minutes, otherwise "Self-test failed!" ap-
pears and the procedure has to be repeated.
If error messages repeatedly appear, contact METTLER TOLEDO Service.
20
Selecting the menu item starts the self-test routine.
The self-test result is displayed after a few seconds.
The meter returns to the system settings menu automatically.
pH/ORP/Ion/Conductivity meter SG78