Picture (B) on Table (A)
D
A
C
A
B
2.) Checking sizes
2.1) Resistance of the protective braid
Actual size:
Ohm
2.2) Residual diversion current
Actual size:
A
2.3) Insulation resistance
Actual size:
MOhm
3.) Functional check
3.1) Functional check of the side railings:
- locking, e.g. play in the locking device
- distances, including under load
- deformation
- wear and tear owing to load
3.2) Functional check of the brakes:
- safety
- braking
- neutral
3.3) Functional check of the drives:
- over the full adjustment range of the motors
(noises, speed, neutral, etc.)
- final switching (e.g. current limiting, main switch)
- if necessary, measure the motor performance first
without load and then with the nominal load (SWL) to
check for wear and tear of the motors
Compare with the values given at the time of
delivery. The manufacturer will provide you with the
limits.
Other:
H
G
F
Page 23
Remark: If the braces of the side
railing are outside (indicated in
dark gray), side railings distance
D
A becomes compulsory for the
lying surface.