Faults, cause, remedy

11
Faults, cause, remedy
Fault
X-zone 5500 cannot be charged.
Battery life is low.
No wireless connection between the
X-zone 5500 devices.
Cable connection is not functioning.
Alarm contact is not switching.
Pump operation is not functioning.
Pump fault during operation.
Flow test has failed.
Horn is too quiet.
66
Cause
Charging unit plug is not in correct
contact with the X-zone 5500.
Inductive charging is not functioning
correctly as the distance between the X-
zone 5500 and the induction charger is
too great.
Battery is not fully charged.
Battery not regularly charged.
Outdoor temperature is too low.
Outdoor temperature is very high.
Battery is faulty.
Inductive charging is not functioning
correctly as the distance between the
X-zone 5500 and the induction charger is
too great.
Wireless function is not activated.
X-zone 5500 devices are very far apart. Position the X-zone 5500 devices more
Wireless connection is made more
difficult by the industrial environment:
e. g. steel walls.
X-zone 5500 devices are covered by
conductive materials (e. g. metal grids).
Network number of the X-zone 5500
device is different.
Wireless frequency of the X-zone 5500 is
different.
Cable plug is not correctly plugged in,
cable assignment is incorrect or cable is
broken.
Cable plug is not correctly plugged in,
cable assignment is incorrect or cable is
broken.
Incorrect holder (diffusion) in position.
Condensate formation with cold and
damp intake air.
Pump outside the specified range.
Flow test has not been performed.
Holder (pump or X-am 5100) incorrectly
attached.
Volume is set too low.
Remedy
Ensure that the charging unit plug is
correctly plugged in.
Check the battery LED.
Check the distance, remove any dirt if
necessary.
Charge battery for at least 14 h - 24 Ah.
Charge battery regularly even when
switched off, at least every 2 months.
Warm up the X-zone 5500.
Use safety housing where appropriate.
Have the battery checked by
DrägerService.
Check the distance, remove any dirt if
necessary.
Activate the wireless function using the
Dräger CC-Vision PC software (see
chapter on page 64).
closely together.
Position additional devices in the chain.
Place X-zone 5500 higher by using a
base, for example (order no. 83 20 645).
Position X-zone 5500 more closely
together.
Place additional devices into the chain.
Connect the X-zone 5500 with a
communication cable (see chapter 5 on
page 55).
Ensure there are no obstructions.
Use X-zone 5500 with the same network
numbers. The network number can be
configured using the Dräger CC-Vision
PC software (see chapter on page 64).
Use X-zone 5500 with the same wireless
frequency.
Check cable connection and
assignment. Ensure that the cable plug
is correctly plugged in.
Check cable connection and
assignment. Ensure that the cable plug
is correctly plugged in.
Position the holder (pump).
Use hose set with water trap (order no.
83 21 527).
Have the pump checked by
DrägerService.
Perform flow test, repeat if necessary.
Reattach holder (pump or X-am 5100)
and ensure the correct position.
Set the volume using the Dräger CC-
Vision PC software (see chapter on
page 64).