Operation With X-Am 5100 (Optional); Commissioning And Performing The Measurement; Pump Test - Dräger X-zone 5500 Instrucciones De Uso

Ocultar thumbs Ver también para X-zone 5500:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Operation with X-am 5100 (optional)

9
Operation with X-am 5100
(optional)
Optionally, the X-zone 5500 with pump (see chapter 20 on
page 76) can be operated with an X-am 5100.
9.1
Commissioning and performing the
measurement
CAUTION
!
Operation with filter only (order no. 83 19 364 und
83 23 016)! Otherwise the pump may be damaged or
the sensor may be heavily soiled.
If no filter is used during pump operation, the warranty
on the pump becomes invalid.
After positioning the holder X-am 5100, a functional
test (see chapter 3.4.3 on page 46) must be conducted
on a successfully completed pump test.
1. Switch on the X-zone 5500 (see" Switching on the
instrument" on page 44) and use the holder X-am 5100.
If the device is in the ON mode, the pump will be
automatically switched on via a switching contact on the
holder X-am 5100.
Pump LED flashes red/green.
2. A pump test will then need to be performed. The pump test
will start automatically.
NOTICE
i
i
The pump test must be performed within 60 seconds,
otherwise a device alarm is issued.
9.2

Pump test

NOTICE
i
i
Dräger recommends inspection of the seals for
damage and for cleanliness in the inside of the cap
before attaching the holder X-am 5100. If the seals are
damaged, the holder X-am 5100 must be replaced. For
cleaning information, see chapter 13 on page 71.
Keep the holder on the X-am 5100 dry and clean when
not in use.
The pump test is performed in the same way as for the
Dräger pump
X-am 1/2/5000. During the pump test, the default
setting acoustic alarm is automatically reduced to
80 dB (A).
64
The pump LED flashes red/
green, accompanied by a
signal tone.
Kink the hose between
the cap and pump inlet for
at least 2 s.
The pump LED lights red.
Unblock the intake flange/
sampling hose.
If the test was successful:
The pump LED lights green,
accompanied by an acoustic
acknowledgment signal.
If the test was not
successful:
The pump LED lights red,
accompanied by a continuous
tone.
The
pump
switches
automatically.
off
Dräger X-zone 5500
00733112.eps
05333112.eps
00833112.eps
00933112.eps
00933112.eps
Tabla de contenido
loading

Tabla de contenido