9.3 Patient Comfort Considerations

Reasons for sensitivity
• Incorrect tip placement. Point should be directed away from
root surfaces.
• Allowing the insert to remain in a static position on the tooth.
Keep the tip in motion and change it's path on the tooth.
• Applying pressure. Use extremely light grasp and pressure,
especially on exposed cementum.
• If sensitivity persists, decrease power setting and/or move from
the sensitive tooth to another and then return.
Section 10: System Care

10.1 Daily Protocol

Start-up procedures at the beginning of the day
1. Switch on the dental unit (refer to dental unit instructions for use).
2. Remove the DENTSPLY Cavitron cable connector assembly
from the handpiece holder and set the power adjustment to 30%
of maximum.
3. Connect Steri-Mate handpiece to the cable connector assembly
(refer to the Infection Control information booklet for sterilizing
instructions).
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through the
handpiece for at least two minutes.
6. Place a sterilized insert into the handpiece using a gentle push-
twist motion.
7. Activate the ultrasonics and adjust the power and lavage control
to your preferred operating positions.
Between patients
1. Remove any ultrasonic inserts and the Steri-Mate handpiece.
Clean and sterilize the handpiece and all inserts used during the
procedure (refer to the Infection Control information booklet and
the Cavitron Ultrasonic Inserts Directions for Use for cleaning
and sterilizing instructions).
2. Clean and disinfect the handpiece cable assembly by applying a
medically approved non-immersion type disinfectant solution,*
carefully following the instructions provided by the solution
manufacturer. To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and connector.
Discard used towel. To disinfect the cable, generously spray
disinfectant on a clean towel and wipe the cable and connector.
Allow the disinfectant solution to air dry.
3. Clean and disinfect all surfaces of the dental treatment control
head according to the manufacturer's instructions.
4. Connect a sterilized Steri-Mate handpiece onto its corresponding
cable connector.
5. Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least thirty seconds.
6. Place a sterilized insert into the handpiece.
ENGLISH • 11