Attach the tubing cap straight to the cannula base. One to two
clicks should be heard.
Rotate the tubing left and right, at least one full turn in each
direction, while gently pulling upward to ensure the tubing cap
is fully engaged and the fluid path is opened.
After proper connection of the tubing to the cannula base, the
newly inserted cannula has to be filled with insulin by priming
the cannula part with the priming volumes indicated below.
Failure to do so results in missed insulin.
Figure A: Disconnection
To disconnect the tubing, gently squeeze tabs on both sides of the
tubing cap and remove it. There is no need to cover the infusion
base as the septum provides closure.
Before you reconnect the tubing cap remove the air bubbles from
the tubing according to the priming instructions of your YpsoPump
Swab septum with a 70 % isopropyl alcohol pad to reduce pos-
sible germ or particle ingress into the fluid path. To reconnect,
follow instructions in figures 12 – 14.
Monitor your blood glucose levels when disconnected and about
2–3 hours after reconnecting.
Priming volume for the tubing with adapter
(volumes are approximate):
45 cm (18 inch): 9 units* (0.09 ml)
60 cm (24 inch): 10 units* (0.10 ml)
80 cm (31 inch): 12 units* (0.12 ml)
110 cm (43 inch): 15 units* (0.15 ml)
Priming volume for the cannula part (head)
(volumes are approximate):
6 mm soft cannula with base: 0.2* units (0.002 ml)
9 mm soft cannula with base: 0.2* units (0.002 ml)
* U-100 Insulin
.
®