•Do not flick the Main Power Switch ON and OFF repeatedly. There is danger that it will blow a fuse.
•Turn off the Control Unit if there are any abnormal vibrations, heat generation, abnormal noise, etc., before or during
use.
•Do not place anything within 10 centimeters around the Control Unit.
•This product should be used only by physicians, dental hygienists, dental technicians and others with professional
training. It should be used only in dental clinics, hospitals or other medical institutions. Assistants should use it under
the supervision of a specializing physician.
•It is recommended that a spare system be kept on hand in case of a breakdown during surgery.
•When operating the product always consider the safety of the patient.
•This product does not consider patient's age (except infants), gender, weight or nationality.
•This product does not consider operator's age (mature person), height, weight, gender, or nationality.
•Do not use this product if the patient has the any of the following conditions:
• Patient complications
• Patient has preexisting condition e.g. (heart, lung, renal, high-blood pressure)
• Patient is Pregnant, patient of child-bearing potential or lactating patient
• Patient is an infant (Age)
•The end user shall be responsible for any judgment that applies this product to a patient.
•Keep the Control Unit on a level surface and make sure it is not subject to strong impact (such as being dropped,
etc.).
•Before using the product, make sure to operate it outside the oral cavity for checking. Then, if you find any
abnormality, immediately stop using it and contact the dealer.
•Disconnect from AC Power Cord and remove the Handpiece Cord after turning off the Control Unit. Pull the plug out
holding the plug itself. Pulling by the Cord will result in damage to the Cord.
•Use only the fuses indicated.
•Do not insert or remove the AC Power Cord or Handpiece Cord with foot on the foot control.
•When the pump gets wet, wipe well and dry it. If the pump remains wet, it may not work normally as the roller inside
the pump may slip.
•If the Irrigation Tube is bent or folded while the Irrigation Pump is in operation, the tube may be damaged or
disconnected.
•If the irrigation seems abnormal, it may be because of wear of the Irrigation Tube or that saline, etc. is leaking from
the tube. Replace the Irrigation Tube.
•The Irrigation Tube cannot be reused. After using, dispose of it as medical waste.
•Do not remove both the Handpiece and cord.
•Do not turn ON the Main Power with the Handpiece disconnected or while foot is on the Foot Control.
•Remove the Tip when inserting or removing the Handpiece Cord or Irrigation Tube. It could cause injuries.
•When in use, always pump cool saline, etc. Insufficient liquid will destroy bone tissue or damage tooth surface.
•In use, never allow the vibrating Tip to touch any prosthesis such as ceramics. Touching might cause break or
chipping.
•In use, never allow the vibrating Tip to touch a metal crown, porcelain crown, etc., which may cause loosening or
breakage, etc.
•Use the Tip within the power range indicated on the tip case. If you use it out of the power range, the Tip might break
or damage an operative site.
•Do not use the Tip that is damaged, bent or rusted. The Tip could break while in use.
•Do not sharpen the Tip or bend it to change the angle. The Tip could break during use or fail to vibrate.
•Even if used within the recommended power range, a Tip could become chipped or break due to an overload. Be
sure to use suction to make sure no pieces of a broken Tip are left in the patient's mouth.
•Make sure that the saline, etc. that emits from the Tip does not come into contact with the control unit. It could cause
color to fade or rusting.
WARNING
CAUTION
• Patient allergic reactions
• Patient has a heart pace maker fitted
2