Case
Water does not come out, or flow
level is low.
Water leak.
Irrigation doesn't stop.
Handpiece LED does not
illuminate. (optic Handpiece)
LCD display does not properly
display.
Part of display is missing.
Settings are not memorized.
Wrong settings are memorized.
* Repairs cannot be made by the customer.
Possible Cause
Foot Control is not connected.
Irrigation Tube is not connected to bottle or
Handpiece.
Irrigation Tube is not properly mounted to
the pump.
The Pump Cover is open.
The Irrigation Tube is broken (leaking).
Leak from the connection between bottle
and Irrigation Tube.
Leak from the connection between
Handpiece and Irrigation Tube.
Leak from Irrigation Tube.
Mode is set to Auto Cleaning.
ON-OFF Switch for Water Flow on Foot
Control is being operated.
Faulty circuit.
LED is burned out.
Failure inside the Control Unit or inside the
Handpiece Cord.
Failure or disconnection of the circuit.
Overheating.
Error Code is displayed.
Failure of the LCD display or driving circuit.
Figures have been temporarily changed.
Foot control is being operated.
An Error Code is displayed.
20
Remedy
Make sure it is connected.
Make sure they are connected.
Make sure it is mounted properly (Refer to
6-5-1 Mounting to the Irrigation Pump).
Close the cover.
Replace the Irrigation Tube (Refer to 6-5
Mounting the Irrigation Tube).
Insert Irrigation Tube Needle all the way into
the bottle (Refer to 6-5-3 Mounting to the
Bottle).
Insert end of Irrigation Tube all the way into
the Handpiece water pipe (Refer to 6-5-2
Mounting to the Handpiece).
Replace the Irrigation Tube (Refer to 6-5
Mounting the Irrigation Tube).
To stop irrigation, press the Flow Key.
Remove foot from the Foot Control.
Contact dealer.*
Contact dealer.*
Contact dealer.*
Contact dealer.*
Heat can cause LCD display malfunction.
Turn the Main Power OFF to cool down.
Refer to the Error Code table.
Turn the Main Power OFF and then back
ON. If part of initial display is still missing,
contact dealer.
Increase or decrease the Program No. by
one, and then return.
Settings cannot be saved while Handpiece
is operating.
Refer to the Error Code table.