7-2 Checking and Setting Up
(1) Checking the Irrigation Tube
Make sure the Irrigation Tube is mounted properly to the bottle, Irrigation Pump, and Handpiece.
(2) Main Power
Turn on the Main Power Switch.
(3) Checking water flow
Open the Tube Clamp that is located between the Irrigation Tube Needle and the Irrigation Pump, press the ON-OFF
Switch for Water Flow on the Foot Control, and make sure that saline comes out of the Tip. With a new tube, it usually
takes several seconds before saline comes out.
(4) Setting the output
CAUTION
Set to the Power indicated on the Tip Case. Use Tips within the appropriate Power Range (Fig. 23). The abbreviations
for the selection modes mean the following : S: SURG mode, E: ENDO mode, P: PERIO mode
1) Switch modes using the Select Mode key.
2) Select a program number using the Program Key.
3) Adjust power level using the Power Key.
4) Adjust flow level using the Flow Key.
* See Fig. 24 for a comparison of outputs of different modes.
CAUTION
Always use a Tip within the mode and power range indicated on the label on the Tip Case. Using a Tip outside its
power range may lead to Tip breakage or damage to tooth surface, gums, or soft tissue.
Fig.23
15
Power OFF Power ON
Fig.24