13.3 Directives and manufacturer's declaration
13.3.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
•
Operation in a professional healthcare facility (e.g. hospital, etc.)
•
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see
Page 44).
The following tables outline the test levels of the tests to be carried out. The higher value applies if
there is a difference between the operating environments listed in the tables below.
Table 4 - casing
Phenomenon
Electrostatic dis
charge
High frequency elec
tromagnetic fields
High frequency elec
tromagnetic fields in
the immediate vicinity
of wireless communic
ation devices
68
Radio ser
Modulation
vice
GSM 1800;
Pulse mod
CDMA 190
0;
GSM 1900;
DECT;
LTE Band
1, 3, 4, 25;
UMTS
Bluetooth
Pulse mod
WLAN 802
.11 b/g/n,
RFID 2450
LTE band 7
WLAN 802
Pulse mod
.11 a/n
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
Maximum
power [W]
2
ulation
217 Hz
2
ulation
217 Hz
0.2
ulation
217 Hz
Interference immunity test level
Professional health
care facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
See Table 9
Distance
Interfer
[m]
ence
immunity
test level
[V/m]
0.3
28
0.3
28
0.3
9
Environments in
areas of home
healthcare
10 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz