Directives And Manufacturer's Declaration; Electromagnetic Environment - Ottobock Kenevo 3C60 Instrucciones De Uso

Ocultar thumbs Ver también para Kenevo 3C60:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 143
Battery charge level
Charger
Battery is charging.
The on time of the LED indicates the current battery charge level.
The on time of the LED gets longer as the charge level increases. It only flashes
briefly at the start of the charging process and stays on continuously at the end of
the charging process.
Battery is fully charged, or the temperature fell above/below the permissible range
during charging. Check current charge level (see Page 50).

13.2.4 Directives and manufacturer's declaration

13.2.4.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see
Page 44).
The following tables outline the test levels of the tests to be carried out. The higher value applies if
there is a difference between the operating environments listed in the tables below.
Table 4 - casing
Phenomenon
Electrostatic dis­
charge
High frequency elec­
tromagnetic fields
High frequency elec­
tromagnetic fields in
the immediate vicinity
of wireless communic­
ation devices
Magnetic fields with
rated power frequen­
cies
Table 5 - alternate current gate for the supply input
Phenomenon
66
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
EMC basic standard
or
test procedure
Interference immunity test level
Professional health­
care facility
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
See Table 9
30 A/m
50 Hz or 60 Hz
Interference immunity test level
Professional health­
care facility
Environments in
areas of home
healthcare
10  V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz
Environments in
areas of home
healthcare