CASE HISTORY OF FAULTS AND THEIR RESOLUTION
Before performing any operation, switch off the appliance and unplug the mains cable from the socket.
PROBLEM
CAUSE
Low battery
Battery not inserted correctly
The device is not
Power cable not correctly plugged into the
working
appliance socket or power socket
- Faulty power supply unit
Collection vessel lid not properly inserted in
No aspiration
the vessel
Aspiration failure caused
Clogged filter
by fluid spillage
Blocked float
Fouling on the float
Vacuum regulator all open
Protection filter blocked
Low and/or no vacuum
Obstructed, kinked or disconnected filter and
power
device connection pipes
Vessel lid overflow valve closed or blocked
Dirty or obstructed or damaged pump
If, after checking the conditions described above, the device still is not working properly, we recommend that you contact your
trusted dealer or an authorised FLAEM service centre nearest to you. You can find a list of all Service Centres at http://www.flaemn-
uova.it/it/info/assistenza
DISPOSAL
Aspiration unit
In accordance with Directive 2012/19/EC, the symbol on the equipment indicates that the equipment to be disposed of (excluding
accessories) is considered as waste and must therefore be subject to 'separate collection' . Therefore, the user must deliver this waste
(or have it delivered) to the separate collection centres set up by the local authorities, or hand it over to the retailer when purchasing a
new appliance of an equivalent type. Separate waste collection and subsequent treatment, recovery and disposal operations promote
the production of equipment from recycled materials and limit the negative environmental and health effects caused by improper waste
management. Unauthorised disposal of the product by the user entails the application of the administrative sanctions provided for in the
laws transposing Directive 2012/19/EC of the member state or country where the product is disposed of.
Accessories (Secretion collection vessel and piping)
Accessories to be disposed of as general waste after a sanitisation cycle.
Packaging
20
Product box
PAP
REMEDY
Recharge the battery
Check the correct insertion of the battery connector
Replace the power supply unit
Insert the lid correctly into the collection vessel
Replace the filter
Remove the lid from the vessel, dismantle the protective
device and remove the float. Then carry out cleaning
operations as per the paragraph on hygienic preparation
Close the regulator completely and check vacuum
power
Replace the filter
Check the condition of the pipes, replace them if they
are obstructed and connect them correctly as per the
'CONNECTION DIAGRAM' in these operating instructions
Unlock the overflow valve, hold the device upright
Take the device to your dealer or an authorised FLAEM
service centre