Usage
10.4
Scanning the image data via
a computer without SmartS-
can
Starting the image plate scanner and soft-
ware
The process of reading out the data is
described here for the imaging software
Vet-Exam Pro.
For further information on using the imag-
ing software, refer to the relevant manual.
1.
Start Vet-Exam Pro.
2.
Select the patient.
3.
Select the corresponding image type in the
menu bar.
4.
Select the device.
5.
Set acquisition mode.
Recording starts directly.
The unit will display an animated visual sym-
bol requesting insertion of the image plate.
Only insert the image plate when the
bar above the animated sequence
has turned to green.
Do not insert any more imaging
plates as long as the animation bar is
blue.
Scanning the image plate
1.
If necessary, read in additional image plates.
When you have finished the last image plate,
click Finish imaging.
30
|
EN
Heed the information on the display.
When inserting the image plate, make
sure that it is assigned to the correct
patient.
2.
Place the light protection cover with the
image plate centrally and straight onto the
plate guide. The opened side of the light
protection cover faces down, the inactive
side of the image plate faces the operator.
The unit automatically detects if the image
plate has been inserted the wrong way
round (active side towards the operator)
and a message to this effect appears on
the display. Turn the image plate over
(inactive side towards the operator) and
re-insert it immediately.
The image plate must not be pushed out of
the light protection cover before it is has
been placed on the plate guide. There is the
risk of image information being erased by
ambient light (see "9 Correct use of image
plates").
3.
Slide the image plate out of its light protec-
tion cover downwards into the device until
the image plate is automatically drawn in.
The light protection cover is held in place by
the plate guide and is not drawn into the
unit.
Make sure that only the image plate, and not
the light protection cover, is inserted into the
unit.
The image data is automatically transmitted
to the imaging software.
After it has been scanned, the image plate is
erased and drops into the collection tray.
4.
Remove the empty light protection cover.
2137100008L02