14. ELECTROMAGNETIC COMPATIBILITY
(EMC) TABLES
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The CEFAR PRIMO PRO Electrotherapy System is intended for use in the electromagnetic environment deified below. The
customer or the user of the CEFAR PRIMO PRO Electrotherapy System should assure that it is used in such an environment
Emissions Tests
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not Applicable - Battery powered
IEC 61000-3-2
Voltage fluctuations
Not Applicable - Battery powered
IEC 61000-3-3
Compliance
The CEFAR PRIMO PRO Electrotherapy System uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
CEFAR PRIMO PRO Electrotherapy System is suitable for
use in all establishments, including domestic establishments
and those directly connected to the public low voltage power
supply network that supplies buildings used for domestic
purposes
Electromagnetic Environment - Guidance
61