Transport
6
Transport
6.1 Safety information
!
IMPORTANT
▸ Take care when handling mattresses to ensure no damage
to the cover. It is recommended that two people lift/carry
mattresses.
▸ Avoid contact with jewellery, nails, abrasive surfaces etc.
▸ Do not drag mattresses.
▸ Avoid contact with wall, door frames, door catches or locks
etc.
▸ Do not transport in roll cages unless completely protected
from the sharp edges of the cage.
6.2 Transport Mode
If it is necessary to move the bed or mattress simply.
1. Turn off power supply.
2. Disconnect control unit power lead (if necessary the air hose).
3. When system is ready to reactivate following
chapter 4.2, page 8.
Air supply hose should be stored by attaching to the fastener at foot of
the mattress.
7
Maintenance
7.1 Inspection
Check mattresses (foam, air insert and cover) for strike-through
(this may include fluid ingress, stains, rips or damage) after the release
of each patient, after ending of the period of use or on a minimum
monthly basis (depending on which occurs first) by a suitably qualified
and competent person.
Check mattresses
1. Unzip the cover completely.
2. Check for any staining on the white underside of the cover.
3. Check for any staining on the interior foam.
4. Replace any stained items and dispose of as per local authority
procedure.
7.2 Cleaning and Care
!
IMPORTANT
All cleaning agents and disinfectants used must be effective,
compatible with one another and must protect the materials
they are used to clean. For further information on decon-
tamination in Healthcare Environments, please refer to 'The
National Institute for Clinical Excellence' guidelines on Infec-
tion Control www.nice.org.uk/CG002 and your local infection
control policy.
Cleaning Handle
1. The exterior of the Handle can be periodically wiped using a cloth
dampened with disinfectant.
Cleaning control unit
WARNING Electrical shock hazard!
▸ Ensure the control unit is disconnected from the mains
electricity supply before cleaning.
▸ Do not spray disinfectant directly on to the control unit, or
immerse the control unit in any type of liquid. This could
result in a severe electrical hazard as this equipment has no
protection against ingress of water.
▸ This equipment is not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
1. Wipe all controls, pump casing and hose fittings with a quaternary
disinfectant solution.