(4) Check electrode positions and ensure that the adhesive electrodes do
not overlap.
(5) Gradually increase the impulse intensity.
(6) The battery is flat; please charge it.
The battery symbol is shown. How to proceed?
Charge the device, following the instructions from section 4.
You are experiencing an unpleasant sensation at the electrodes. How
to proceed?
(1) The electrodes are not positioned correctly. Check their positions and
re-position, if necessary.
(2) The gel pads are worn. This may cause irritated skin, as even distribu-
tion of the current across the entire area is no longer guaranteed. For
this reason, the electrodes should be replaced.
Skin in the treatment area turns red. How to proceed?
Immediately stop treatment and wait until your skin has returned to its
normal condition. If the redness is under the electrode and disappears
quickly, there is no risk – this is caused by the locally stimulated, in-
creased blood flow.
However, consult your doctor before you continue treatment if the skin
irritation persists and if it is accompanied by an itchy sensation or in-
flammation. This may be caused by an allergic reaction to the adhesive
surface.
The device is getting too hot. How to proceed?
Switch to the lower heat level or switch the heat function off entirely.
16. REPLACEMENT PARTS AND WEARING
PARTS
You can obtain the following replacement parts directly from Customer
Services:
Designation
8 x gel pads (45 x 45 mm)
Item number and/or order number
Item 646.55
17. TECHNICAL SPECIFICATIONS
Type
Output waveform
Pulse length
Pulse frequency
Output voltage
Output current
Voltage supply
Treatment time
Intensity
Heating levels
Operating conditions
Storage conditions
transport conditions:
Dimensions
Weight
Height limit for use
Maximum permissible
atmospheric pressure:
The serial number is located on the device.
Note: If the device is not used according to the instructions specified,
perfect functionality cannot be guaranteed!
37
EM 59
Biphasic rectangular pulses
50–450 µs
1–150 Hz
Max. 100 Vpp (500 ohm)
Max. 200 mApp (500 ohm)
Lithium-ion rechargeable battery,
2000mAh, 3.7V
Adjustable from 5 to 100 minutes
Adjustable from 0 to 50
low (41 °C) ; high (43 °C)
5°C–40°C (41°F–104°F) at a relative air
humidity of 40–70%
0°C–40°C (32°F–104°F) at a relative air
humidity of 0–90%
-25°C- 70°C (-13°F-158°F) at a relative air
humidity of 0-90%
Approx. 139 x 66 x 26 mm (including belt
clip)
Approx. 125 g (including belt clip),
3,000 m
700–1,060 hPa