PRECAUTIONS:
• Always pull firmly on electrodes to remove them from the skin to pre-
vent injuries in the unusual case of highly sensitive skin.
• Keep the device away from sources of heat (heating devices such as
driers or ovens) and do not use it in close proximity (approx. 1 m) to
shortwave or microwave devices (e.g. mobile phones), as doing so
can result in unpleasant current peaks.
• Do not expose the device to direct sunlight or high temperatures.
• Protect the device from dust, dirt and humidity.
• Never immerse the device in water or other liquids.
• The device is suitable for self-treatment.
• For hygiene reasons, the electrodes may only be used on one person.
• If the device does not work properly, or if you feel unwell or experi-
ence pain, stop using it immediately.
• Switch off the unit or the respective channel first before removing or
relocating electrodes to prevent unintentional stimulation.
• Do not modify electrodes (e.g. by cutting them), as this increases the
current density, which is potentially hazardous (max. recommended
output value for the electrodes is 9 mA/cm², an effective current den-
sity beyond 2 mA/cm² requires increased awareness).
• Do not use the device whilst asleep, driving a vehicle or operating
machinery.
• Do not use whilst undertaking any activity where an unexpected re-
action (e.g. strong muscle contractions even at low intensity) could
be dangerous.
• Ensure that no metallic objects (e.g. belt buckles or necklaces) come
into contact with the electrodes during stimulation. If you are wearing
jewellery or have piercings in the area to be treated (e.g. a navel pierc-
ing), these must be removed before using the device. Failure to do so
could result in spot burns.
• Keep the device away from children.
• Make sure not to confuse the electrode cables including contacts
with your headphones or other devices, and do not connect the elec-
trodes to other devices.
• Do not use the device whilst using other devices that transmit electri-
cal impulses to your body.
• Do not use in the vicinity of highly flammable substances, gases or
explosives.
• The actual temperature may vary depending on the condition of your
skin, your age, the location of the pain, etc.
• If the heat function feels too hot, stop treatment immediately. You
can continue the TENS, EMS or massage treatment without the heat
function.
• During the initial few minutes, use the device while sitting or lying
down to minimise the risk of injury resulting from isolated cases of
vagal responses (feeling of faintness). If you feel faint, immediately
switch off the device, lie down and support your legs in an elevated
position (approx. 5 –10 min.).
• Treatment of the skin with moisturising lotions or ointments before-
hand is not recommended as this considerably increases the gel pad
wear and may cause unpleasant current peaks.
• This device is not intended for use by children or people with restrict-
ed physical, sensory (e.g. reduced sensitivity to pain) or mental skills
or a lack of experience and/or lack of knowledge, unless they are
supervised by a person who is responsible for their safety or who are
instructed by such a person in how to use the device.
• If the adhesive capability of the gel pads decreases, please replace
them immediately. Do not use the device until you have the new gel
pads. Otherwise the unequal adhesion of the gel pads may lead to
skin injuries. Replace the gel pads with new ones at the latest after
having used them 20 times.
Damage
• If the device is damaged, do not use it and contact your retailer or the
specified Customer Services address.
• To ensure that the device functions effectively, do not drop it or dis-
mantle it.
• Check the device for signs of wear and tear or damage. If there are
such signs of wear and tear or damage or if the device was used
25