Dräger Vista 120 Anexo página 3

Ocultar thumbs Ver también para Vista 120:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 7
Supplement to the Instructions for Use
Vista 120 Part no. 2680051, Edition: 4 – 2014-12 and higher
This supplement is used for Vista 120 Software 2.21.
 Keep this supplement with the instructions for use of the medical device.
 This supplement updates the information given in the following chapters of the instructions
for use.
Overview
Abbreviations
Delete the following information: PAWP Pulmonary artery wedge pressure.
Getting started
Connecting the power cord
Delete "3-phase".
Alarms
Selecting alarm tone type
The user can select the alarm tone type as desired.
1. Select Menu > Maintenance > User Maintain, and enter the required password ABC.
2. Select Alarm Setup, and set Alarm Tone to Standard or Mode 1.
 Standard: Standard alarm sound according to IEC 60601-1-8.
 Mode 1: User customized alarm sound according to clinical applications.
User interface
Changing Parameter and Waveform Colors
To change the parameter and waveform color:
1. Select Menu > Maintenance > User Maintain, and enter the required password ABC.
2. Select Color Setup to make color changes on parameter and waveform.
Supplement to the Instructions for Use Vista 120 SW 2.21
English
3
loading