Set the electrode gain as low as possible in order to reduce interference from powerful elec
tromagnetic radiation (e.g. visible or or concealed theft prevention systems at the
entrance/exit of stores), metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong electromagnetic interference (e.g. high-voltage lines, transmitters, trans
former stations, computer tomographs, magnetic resonance tomographs, etc.).
CAUTION
Use of unsuitable prosthetic components
Injury due to unexpected product behaviour.
►
Use the product only in combination with components listed in the section "Combination
Possibilities" (see Page 15).
4.5 Information on Use
CAUTION
Risk of pinching between the prehensile arms/fingertips
Injury due to pinching of body parts.
►
Ensure that no body parts are between the prehensile arms/fingertips when using the
product.
CAUTION
Mechanical stress on the product
Injury due to faulty control or malfunction of the product.
►
Do not subject the product to mechanical vibrations or impacts.
►
Check the product for visible damage before each use.
CAUTION
Operation outside of the permissible temperature range
Injury due to faulty control or malfunction of the product.
►
Avoid operation in areas outside the permissible temperature range (see Page 23).
CAUTION
Penetration of dirt and humidity into the product
Injury due to unexpected behaviour of the prosthesis system.
►
Ensure that neither solid particles nor liquids can penetrate into the product.
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
►
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der
maClean).
►
Use only the following products for cleaning/disinfecting the inner socket:
Cleaning: 453H10=1 Ottobock DermaClean
Disinfection: 453H16 Ottobock DermaDesinfect.
17